40 CFR 86.125-94
§ 86.125-94 Methane analyzer calibration.
United States · 40 CFR — Protection of Environment · Status: effective
Cite this
- Citation
- 40 CFR 86.125-94, § 86.125-94 Methane analyzer calibration, United States, version 1 as recorded 2026-07-09, yourstate.us, https://yourstate.us/provision/179666
- Permanent ID
ys:prov:179666@1- SHA-256
74cc9f4271f2e815b71f10dd2f87ecbb80ea243a5d54ea99125f0bcb7b5caf04
The hash is SHA-256 of this version's text, with every run of whitespace collapsed to a single space and the ends trimmed. The ID always leads back here, and checking it says whether the text you cited is still the current version.
Full text
Prior to introduction into service and monthly thereafter, the methane analyzer shall be calibrated:
(a) Follow the manufacturer's instructions for instrument startup and operation. Adjust the analyzer to optimize performance.
(b) Zero the methane analyzer with zero-grade air.
(c) Calibrate on each normally used operating range with CH4 in air with nominal concentrations of 15, 30, 45, 60, 75, and 90 percent of that range. Additional calibration points may be generated. For each range calibrated, if the deviation from a least-squares best-fit straight line is 2 percent or less of the value at each data point, concentration values may be calculated by use of a single calibration factor for that range. If the deviation exceeds 2 percent at any point, the best-fit non-linear equation which represents the data to within 2 percent of each test point shall be used to determine concentration.