40 CFR 1065.366
§ 1065.366 Interference verification for FTIR analyzers.
United States · 40 CFR — Protection of Environment · Status: effective
Cite this
- Citation
- 40 CFR 1065.366, § 1065.366 Interference verification for FTIR analyzers, United States, version 1 as recorded 2026-07-09, yourstate.us, https://yourstate.us/provision/195745
- Permanent ID
ys:prov:195745@1- SHA-256
aae75e8b5dc32fff28ddc15063e04a83ca62ede1bf0ec5a0e5eca4fff85b30f3
The hash is SHA-256 of this version's text, with every run of whitespace collapsed to a single space and the ends trimmed. The ID always leads back here, and checking it says whether the text you cited is still the current version.
Full text
(a) Scope and frequency. If you measure CH4, C2H6, NMHC, or NMNEHC using an FTIR analyzer, verify the amount of interference after initial analyzer installation and after major maintenance.
(b) Measurement principles. Certain species can interfere with analyzers by causing a response similar to the target analyte. If the analyzer uses compensation algorithms that utilize measurements of other gases to meet this interference verification, a correct result depends on simultaneously conducting these other measurements to test the compensation algorithms during the analyzer interference verification.
(c) System requirements. An FTIR analyzer must have combined interference that is within ±2% of the flow-weighted mean concentration of CH4, NMHC, or NMNEHC expected at the standard, though we strongly recommend a lower interference that is within ±1%.
(d) Procedure. Perform the interference verification for an FTIR analyzer using the same procedure that applies for N2O analyzers in § 1065.375(d).