§ 84.91 Breathing resistance test; exhalation.
United States · 42 CFR — Public Health · Status: effective
Cite this
- Citation
- 42 CFR 84.91, § 84.91 Breathing resistance test; exhalation, United States, version 1 as recorded 2026-07-09, yourstate.us, https://yourstate.us/provision/201708
- Permanent ID
ys:prov:201708@1- SHA-256
3f5d1f28345865fddc1b41f1246bb85fa0de704506627d9bd20d530c9837dcf1
The hash is SHA-256 of this version's text, with every run of whitespace collapsed to a single space and the ends trimmed. The ID always leads back here, and checking it says whether the text you cited is still the current version.
Full text
(a) Resistance to exhalation airflow will be measured in the facepiece or mouthpiece of open-circuit apparatus with air flowing at a continuous rate of 85 liters per minute.
(b) The exhalation resistance of demand apparatus shall not exceed 25 mm. (1 inch) water-column height.
(c) The exhalation resistance of pressure-demand apparatus shall not exceed the static pressure in the facepiece by more than 51 mm. (2 inches) water-column height.
(d) The static pressure (at zero flow) in the facepiece shall not exceed 38 mm. (1.5 inches) water-column height.
(e) Resistance to exhalation airflow will be measured in the facepiece or mouthpiece of closed-circuit apparatus with a breathing machine as described in § 84.88, and the exhalation resistance shall not exceed 51 mm. (2 inches) water-column height.