42 CFR 84.94
§ 84.94 Gas flow test; closed-circuit apparatus.
United States · 42 CFR — Public Health · Status: effective
Cite this
- Citation
- 42 CFR 84.94, § 84.94 Gas flow test; closed-circuit apparatus, United States, version 1 as recorded 2026-07-09, yourstate.us, https://yourstate.us/provision/201711
- Permanent ID
ys:prov:201711@1- SHA-256
adec90226bfd63d3ae63b01e8a7af835d30ca82d12620050302f6e87e7f9b1b3
The hash is SHA-256 of this version's text, with every run of whitespace collapsed to a single space and the ends trimmed. The ID always leads back here, and checking it says whether the text you cited is still the current version.
Full text
(a) Where oxygen is supplied by a constant-flow device only, the rate of flow shall be at least 3 liters per minute for the entire rated service time of the apparatus.
(b) Where constant flow is used in conjunction with demand flow, the constant flow shall be greater than 1.5 liters per minute for the entire rated service time.
(c) All demand-flow devices shall provide at least 30 liters of oxygen per minute when in the fully open position.