§ 84.148 Type C supplied-air respirator, continuous flow class; minimum requirements.

United States · 42 CFR — Public Health · Status: effective

Get this as JSONEmbed this
Cite this
Citation
42 CFR 84.148, § 84.148 Type C supplied-air respirator, continuous flow class; minimum requirements, United States, version 1 as recorded 2026-07-09, yourstate.us, https://yourstate.us/provision/201761
Permanent ID
ys:prov:201761@1
SHA-256
4e0b0ba66e08ecda81492e8222332830d4684637f07384d31d0646711edf42aa

The hash is SHA-256 of this version's text, with every run of whitespace collapsed to a single space and the ends trimmed. The ID always leads back here, and checking it says whether the text you cited is still the current version.

Full text

(a) Respirators tested under this section shall be approved only when they supply respirable air at the pressures and quantities required. (b) The pressure at the inlet of the hose connection shall not exceed 863 kN/m. 2 (125 pounds per square inch gage). (c) Where the pressure at any point in the supply system exceeds 863 kN/m. 2 (125 pounds per square inch gage), the respirator shall be equipped with a pressure-release mechanism that will prevent the pressure at the hose connection from exceeding 863 kN/m. 2 (125 pounds per square inch gage) under any conditions.