§ 84.153 Airflow resistance test, Type A and Type AE supplied-air respirators; minimum requirements.

United States · 42 CFR — Public Health · Status: effective

Get this as JSONEmbed this
Cite this
Citation
42 CFR 84.153, § 84.153 Airflow resistance test, Type A and Type AE supplied-air respirators; minimum requirements, United States, version 1 as recorded 2026-07-09, yourstate.us, https://yourstate.us/provision/201766
Permanent ID
ys:prov:201766@1
SHA-256
e38a73f4c307978a0c85a3c1d403827e9ed36d43fcb02561e8e96f9b56d69fb6

The hash is SHA-256 of this version's text, with every run of whitespace collapsed to a single space and the ends trimmed. The ID always leads back here, and checking it says whether the text you cited is still the current version.

Full text

(a) Airflow resistance will be determined when the respirator is completely assembled with the respiratory-inlet covering, the air-supply device, and the maximum length of air-supply hose coiled for one-half its length in loops 1.5 to 2.1 m. (5 to 7 feet) in diameter. (b) The inhalation resistance, drawn at the rate of 85 liters (3 cubic feet) per minute when the blower is not operating or under any practical condition of blower operation shall not exceed the following amounts: (c) The exhalation resistance shall not exceed 25 mm. (1 inch) of water-column height at a flow rate of 85 liters (3 cubic feet) per minute when the blower is not operating or under any practical condition of blower operation.