§ 84.154 Airflow resistance test; Type B and Type BE supplied-air respirators; minimum requirements.
United States · 42 CFR — Public Health · Status: effective
Cite this
- Citation
- 42 CFR 84.154, § 84.154 Airflow resistance test; Type B and Type BE supplied-air respirators; minimum requirements, United States, version 1 as recorded 2026-07-09, yourstate.us, https://yourstate.us/provision/201767
- Permanent ID
ys:prov:201767@1- SHA-256
32746d653e1e93e10de42a7dafc632c2bc1f4f44f6facf5e0e5d92a9d6b0542d
The hash is SHA-256 of this version's text, with every run of whitespace collapsed to a single space and the ends trimmed. The ID always leads back here, and checking it says whether the text you cited is still the current version.
Full text
(a) Airflow resistance shall be determined when the respirator is completely assembled with the respiratory-inlet covering and the hose in the maximum length to be considered for approval, coiled in loops 1.5 to 2.1 m. (5 to 7 feet) in diameter.
(b) Airflow resistance shall not exceed 38 mm. (1.5 inches) of water-column height to air drawn at the flow rate of 85 liters (3 cubic feet) per minute.
(c) The exhalation resistance shall not exceed 25 mm. (1 inch) of water-column height at this flow rate.