§ 84.158 Exhalation valve leakage test.

United States · 42 CFR — Public Health · Status: effective

Get this as JSONEmbed this
Cite this
Citation
42 CFR 84.158, § 84.158 Exhalation valve leakage test, United States, version 1 as recorded 2026-07-09, yourstate.us, https://yourstate.us/provision/201771
Permanent ID
ys:prov:201771@1
SHA-256
b6f1ecdd071a203512c0c0a3cda12838d78fca6e5c1d4482352874e472b202db

The hash is SHA-256 of this version's text, with every run of whitespace collapsed to a single space and the ends trimmed. The ID always leads back here, and checking it says whether the text you cited is still the current version.

Full text

(a) Dry exhalation valves and valve seats will be subjected to a suction of 25 mm. water-column height while in a normal operating position. (b) Leakage between the valve and valve seat shall not exceed 30 milliliters per minute.