42 CFR 84.175
§ 84.175 Instantaneous filter efficiency level determination test—PAPR series HE, PAPR100-N, and PAPR100-P filtration.
United States · 42 CFR — Public Health · Status: effective
Cite this
- Citation
- 42 CFR 84.175, § 84.175 Instantaneous filter efficiency level determination test—PAPR series HE, PAPR100-N, and PAPR100-P filtration, United States, version 1 as recorded 2026-07-09, yourstate.us, https://yourstate.us/provision/201784
- Permanent ID
ys:prov:201784@1- SHA-256
4afe668e0a96fe3cac1e7155537e957cdbd0d94d74daa84dece342a8d19b0100
The hash is SHA-256 of this version's text, with every run of whitespace collapsed to a single space and the ends trimmed. The ID always leads back here, and checking it says whether the text you cited is still the current version.
Full text
(a) Three filters from each powered air-purifying particulate respirator for efficiency will be tested against a neat cold-nebulized dioctyl phthalate (DOP) or equivalent aerosol at 25 ± 5 °C that has been neutralized to the Boltzmann equilibrium state.
(b) Single air-purifying particulate respirator filter units will be tested in an atmosphere concentration of 100 mg/m 3 of DOP at the following continuous flow rates for a period of 5 to 10 seconds:
(c) Powered air-purifying particulate respirators with multiple filter units will be tested by dividing the flow rate specified in paragraph (b) of this section by the total number of filters used.
(d) The filter will be mounted on a connector in the same manner as used on the respirator and the total efficiency must be ≥99.97 percent.
Legislative history