§ 84.200 Inhalation and exhalation valves; minimum requirements.
United States · 42 CFR — Public Health · Status: effective
Cite this
- Citation
- 42 CFR 84.200, § 84.200 Inhalation and exhalation valves; minimum requirements, United States, version 1 as recorded 2026-07-09, yourstate.us, https://yourstate.us/provision/201802
- Permanent ID
ys:prov:201802@1- SHA-256
73fa667b94ca1289ad6b6db84822f2d97c66d606162d619f4acfa5eaa44faf3b
The hash is SHA-256 of this version's text, with every run of whitespace collapsed to a single space and the ends trimmed. The ID always leads back here, and checking it says whether the text you cited is still the current version.
Full text
(a) Inhalation and exhalation valves shall be provided where necessary and protected against distortion.
(b) Inhalation valves shall be designed and constructed to prevent excessive exhaled air from entering cartridges or adversely affecting canisters.
(c) Exhalation valves shall be—
(1) Protected against damage and external influence; and
(2) Designed and constructed to prevent inward leakage of contaminated air.