42 CFR 84.207
§ 84.207 Bench tests; gas and vapor tests; minimum requirements; general.
United States · 42 CFR — Public Health · Status: effective
Cite this
- Citation
- 42 CFR 84.207, § 84.207 Bench tests; gas and vapor tests; minimum requirements; general, United States, version 1 as recorded 2026-07-09, yourstate.us, https://yourstate.us/provision/201809
- Permanent ID
ys:prov:201809@1- SHA-256
539635347a5bba2cddc185a3c91bdd7af31a864dbf3559b9d908295d4cc912e0
The hash is SHA-256 of this version's text, with every run of whitespace collapsed to a single space and the ends trimmed. The ID always leads back here, and checking it says whether the text you cited is still the current version.
Full text
(a) Bench tests will be made on an apparatus that allows the test atmosphere at 50 ±5 percent relative humidity and room temperature, approximately 25 °C, to enter the cartridges continuously at predetermined concentrations and rates of flow, and that has means for determining the test life of the cartridges.
(b) Where two cartridges are used in parallel on a chemical cartridge respirator, the bench test will be performed with the cartridges arranged in parallel, and the test requirements will apply to the combination rather than to the individual cartridges.
(c) Three cartridges or pairs of cartridges will be removed from containers and tested as received from the applicant.
(d) Two air purifying cartridges or pairs of cartridges will be equilibrated at room temperature by passing 25 percent relative humidity air through them at the flow rate of 25 liters per minute (l.p.m.) for 6 hours.
(e) Two air purifying cartridges or pairs of cartridges will be equilibrated by passing 85 percent relative humidity air through them at the flow rate of 25 l.p.m.
(f) All cartridges will be resealed, kept in an upright position, at room temperatures, and tested within 18 hours.
(g) Cartridges will be tested and shall meet the minimum requirements set forth in Table 11 of this subpart.
(h) Powered air-purifying respirators with a cartridge(s) and particulate filter(s) must meet the as-received minimum service-life requirements and half of the equilibrated minimum service-life requirements set forth in table 11 of subpart L using the flows specified in subpart K, § 84.175(b) and equilibrated in accordance with paragraphs (a) through (g) of this section using the flows specified in subpart K, § 84.175(b).