42 CFR 84.307
§ 84.307 Environmental treatments.
United States · 42 CFR — Public Health · Status: effective
Cite this
- Citation
- 42 CFR 84.307, § 84.307 Environmental treatments, United States, version 1 as recorded 2026-07-09, yourstate.us, https://yourstate.us/provision/201829
- Permanent ID
ys:prov:201829@1- SHA-256
48f292b1b2f186caf9009a4bb003fde1ba71363428c6b786594d435d00c6c58e
The hash is SHA-256 of this version's text, with every run of whitespace collapsed to a single space and the ends trimmed. The ID always leads back here, and checking it says whether the text you cited is still the current version.
Full text
(a) Four units submitted for approval will be tested for capacity and performance, pursuant to the requirements of §§ 84.303 through 84.305, after exposure to environmental treatments simulating extreme storage temperatures, shock, and vibration.
(b) The units will be stored for 16 hours at a temperature of −45 °C and for 48 hours at a temperature of 71 °C. Units will be returned to room temperature between high and low temperature treatments. The maximum rate of change for thermal loading shall not exceed 3 °C per minute and constant temperatures shall be maintained within ±2 °C.
(c) The units, in the casing in which they are deployed for individual use, will be subjected to physical shock according to the following procedure:
(1) The unit will be dropped six times from a height of 1 meter onto a concrete surface; and
(2) Each drop will test a different orientation of the unit, with two drops along each of its three major axes (top to bottom, left to right, and front to back).
(d) The units will be subjected to vibration according to the following procedure:
(1) The unit will be firmly secured to a shaker table, which will be vibrated with motion applied along a single axis for 180 minutes;
(2) The unit will be vibrated one axis at a time along each of three axes for a total of 9 hours; and
(3) The vibration frequency regimen applied to each axis will be cyclical, repeating the sequence and specifications provided in Table 5 every 20 minutes.
Table 5—Vibration Test Sequence