43 CFR 3175.132
§ 3175.132 Testing of reference accuracy.
United States · 43 CFR — Public Lands: Interior · Status: effective
Cite this
- Citation
- 43 CFR 3175.132, § 3175.132 Testing of reference accuracy, United States, version 1 as recorded 2026-07-09, yourstate.us, https://yourstate.us/provision/212298
- Permanent ID
ys:prov:212298@1- SHA-256
8c2127f61c4df9c868049db998da6d99a1402f0f154669891314346bc74f6bc8
The hash is SHA-256 of this version's text, with every run of whitespace collapsed to a single space and the ends trimmed. The ID always leads back here, and checking it says whether the text you cited is still the current version.
Full text
(a) The following reference test conditions must be maintained for the duration of the testing:
(1) Ambient air temperature must be between 59 °F and 77 °F and must not vary over the duration of the test by more than ±2 °F;
(2) Relative humidity must be between 45 percent and 75 percent and must not vary over the duration of the test by more than ±5 percent;
(3) Atmospheric pressure must be between 12.46 psi and 15.36 psi and must not vary over the duration of the test by more than ±0.2 psi;
(4) The transducer must be isolated from any externally induced vibrations;
(5) The transducer must be mounted according to the manufacturer's specifications in the same manner as it would be mounted in normal field operations;
(6) The transducer must be isolated from any external electromagnetic fields; and
(7) For reference accuracy testing of differential-pressure transducers, the downstream side of the transducer must be vented to the atmosphere.
(b) Before reference testing begins, the following pre-conditioning steps must be followed:
(1) After power is applied to the transducer, it must be allowed to stabilize for at least 30 minutes before applying any input pressure or temperature;
(2) The transducer must be exercised by applying three full-range traverses in each direction; and
(3) The transducer must be calibrated according to manufacturer specifications if a calibration is required or recommended by the manufacturer.
(c) Immediately following preconditioning, the transducer must be tested at least three times for accuracy under § 3175.131(h). The results of these tests must be used to determine the transducer's reference accuracy under § 3175.135.
Legislative history
This is a federal regulation, adopted through agency rulemaking under the Administrative Procedure Act — not enacted by a recorded vote of Congress.