§ 35.432 Calibration measurements of brachytherapy sources.

United States · 10 CFR — Energy · Status: effective

Get this as JSONEmbed this
Cite this
Citation
10 CFR 35.432, § 35.432 Calibration measurements of brachytherapy sources, United States, version 1 as recorded 2026-07-09, yourstate.us, https://yourstate.us/provision/36988
Permanent ID
ys:prov:36988@1
SHA-256
9b81656347aedda3151f73bd8f955f2efd8cebf4ced1e68a5dcee505468618c8

The hash is SHA-256 of this version's text, with every run of whitespace collapsed to a single space and the ends trimmed. The ID always leads back here, and checking it says whether the text you cited is still the current version.

Full text

(a) Before the first medical use of a brachytherapy source on or after October 24, 2002, a licensee shall have— (1) Determined the source output or activity using a dosimetry system that meets the requirements of § 35.630(a); (2) Determined source positioning accuracy within applicators; and (3) Used published protocols currently accepted by nationally recognized bodies to meet the requirements of paragraphs (a)(1) and (a)(2) of this section. (b) Instead of a licensee making its own measurements as required in paragraph (a) of this section, the licensee may use measurements provided by the source manufacturer or by a calibration laboratory accredited by the American Association of Physicists in Medicine that are made in accordance with paragraph (a) of this section. (c) A licensee shall mathematically correct the outputs or activities determined in paragraph (a) of this section for physical decay at intervals consistent with 1 percent physical decay. (d) A licensee shall retain a record of each calibration in accordance with § 35.2432.

Legislative history

This is a federal regulation, adopted through agency rulemaking under the Administrative Procedure Act — not enacted by a recorded vote of Congress.