§ 35.457 Therapy-related computer systems.
United States · 10 CFR — Energy · Status: effective
Cite this
- Citation
- 10 CFR 35.457, § 35.457 Therapy-related computer systems, United States, version 1 as recorded 2026-07-09, yourstate.us, https://yourstate.us/provision/36990
- Permanent ID
ys:prov:36990@1- SHA-256
09acab8173351782f23bff56bd75874fc719820cdd325b84ba00db5152d6ceea
The hash is SHA-256 of this version's text, with every run of whitespace collapsed to a single space and the ends trimmed. The ID always leads back here, and checking it says whether the text you cited is still the current version.
Full text
The licensee shall perform acceptance testing on the treatment planning system of therapy-related computer systems in accordance with published protocols accepted by nationally recognized bodies. At a minimum, the acceptance testing must include, as applicable, verification of:
(a) The source-specific input parameters required by the dose calculation algorithm;
(b) The accuracy of dose, dwell time, and treatment time calculations at representative points;
(c) The accuracy of isodose plots and graphic displays; and
(d) The accuracy of the software used to determine sealed source positions from radiographic images.