§ 35.643 Periodic spot-checks for remote afterloader units.
United States · 10 CFR — Energy · Status: effective
Cite this
- Citation
- 10 CFR 35.643, § 35.643 Periodic spot-checks for remote afterloader units, United States, version 1 as recorded 2026-07-09, yourstate.us, https://yourstate.us/provision/37007
- Permanent ID
ys:prov:37007@1- SHA-256
5579825e852ae27d6cbeb01410d560b6b208011f4d7f7c65faffd3a4f9486398
The hash is SHA-256 of this version's text, with every run of whitespace collapsed to a single space and the ends trimmed. The ID always leads back here, and checking it says whether the text you cited is still the current version.
Full text
(a) A licensee authorized to use a remote afterloader unit for medical use shall perform spot-checks of each remote afterloader facility and on each unit—
(1) Before the first use of a high dose-rate, medium dose-rate, or pulsed dose-rate remote afterloader unit on a given day;
(2) Before each patient treatment with a low dose-rate remote afterloader unit; and
(3) After each source installation.
(b) A licensee shall perform the measurements required by paragraph (a) of this section in accordance with written procedures established by the authorized medical physicist. That individual need not actually perform the spot check measurements.
(c) A licensee shall have the authorized medical physicist review the results of each spot-check within 15 days. The authorized medical physicist shall notify the licensee as soon as possible in writing of the results of each spot-check.
(d) To satisfy the requirements of paragraph (a) of this section, spot-checks must, at a minimum, assure proper operation of—
(1) Electrical interlocks at each remote afterloader unit room entrance;
(2) Source exposure indicator lights on the remote afterloader unit, on the control console, and in the facility;
(3) Viewing and intercom systems in each high dose-rate, medium dose-rate, and pulsed dose-rate remote afterloader facility;
(4) Emergency response equipment;
(5) Radiation monitors used to indicate the source position;
(6) Timer accuracy;
(7) Clock (date and time) in the unit's computer; and
(8) Decayed source(s) activity in the unit's computer.
(e) If the results of the checks required in paragraph (d) of this section indicate the malfunction of any system, a licensee shall lock the control console in the off position and not use the unit except as may be necessary to repair, replace, or check the malfunctioning system.
(f) A licensee shall retain a record of each check required by paragraph (d) of this section and a copy of the procedures required by paragraph (b) of this section in accordance with § 35.2643.