§ 35.657 Therapy-related computer systems.
United States · 10 CFR — Energy · Status: effective
Cite this
- Citation
- 10 CFR 35.657, § 35.657 Therapy-related computer systems, United States, version 1 as recorded 2026-07-09, yourstate.us, https://yourstate.us/provision/37012
- Permanent ID
ys:prov:37012@1- SHA-256
979d621d209ceeda09a7dd84940ba7a9d13549ed71ad2acd3218f98936b4117a
The hash is SHA-256 of this version's text, with every run of whitespace collapsed to a single space and the ends trimmed. The ID always leads back here, and checking it says whether the text you cited is still the current version.
Full text
The licensee shall perform acceptance testing on the treatment planning system of therapy-related computer systems in accordance with published protocols accepted by nationally recognized bodies. At a minimum, the acceptance testing must include, as applicable, verification of:
(a) The source-specific input parameters required by the dose calculation algorithm;
(b) The accuracy of dose, dwell time, and treatment time calculations at representative points;
(c) The accuracy of isodose plots and graphic displays;
(d) The accuracy of the software used to determine sealed source positions from radiographic images; and
(e) The accuracy of electronic transfer of the treatment delivery parameters to the treatment delivery unit from the treatment planning system.