yourstate.us
10 CFR 35.2632

§ 35.2632 Records of teletherapy, remote afterloader, and gamma stereotactic radiosurgery full calibrations.

United States · 10 CFR — Energy · Status: effective

Get this as JSONEmbed this
Cite this
Citation
10 CFR 35.2632, § 35.2632 Records of teletherapy, remote afterloader, and gamma stereotactic radiosurgery full calibrations, United States, version 1 as recorded 2026-07-09, yourstate.us, https://yourstate.us/provision/37039
Permanent ID
ys:prov:37039@1
SHA-256
d0328c8300e6b53909d85636f512d02f861ac99eb7db009e0a33ad4370ae3110

The hash is SHA-256 of this version's text, with every run of whitespace collapsed to a single space and the ends trimmed. The ID always leads back here, and checking it says whether the text you cited is still the current version.

Full text

(a) A licensee shall maintain a record of the teletherapy unit, remote afterloader unit, and gamma stereotactic radiosurgery unit full calibrations required by §§ 35.632, 35.633, and 35.635 for 3 years. (b) The record must include— (1) The date of the calibration; (2) The manufacturer's name, model number, and serial number of the teletherapy, remote afterloader, and gamma stereotactic radiosurgery unit(s), the source(s), and the instruments used to calibrate the unit(s); (3) The results and an assessment of the full calibrations; (4) The results of the autoradiograph required for low dose-rate remote afterloader units; and (5) The signature of the authorized medical physicist who performed the full calibration.

Legislative history

This is a federal regulation, adopted through agency rulemaking under the Administrative Procedure Act — not enacted by a recorded vote of Congress.