10 CFR 35.2643
§ 35.2643 Records of periodic spot-checks for remote afterloader units.
United States · 10 CFR — Energy · Status: effective
Cite this
- Citation
- 10 CFR 35.2643, § 35.2643 Records of periodic spot-checks for remote afterloader units, United States, version 1 as recorded 2026-07-09, yourstate.us, https://yourstate.us/provision/37041
- Permanent ID
ys:prov:37041@1- SHA-256
0c3b403040491cc0800cc61b130419d816bef09fe2e038ef414304023bfcffa5
The hash is SHA-256 of this version's text, with every run of whitespace collapsed to a single space and the ends trimmed. The ID always leads back here, and checking it says whether the text you cited is still the current version.
Full text
(a) A licensee shall retain a record of each spot-check for remote afterloader units required by § 35.643 for 3 years.
(b) The record must include, as applicable—
(1) The date of the spot-check;
(2) The manufacturer's name, model number, and serial number for the remote afterloader unit and source;
(3) An assessment of timer accuracy;
(4) Notations indicating the operability of each entrance door electrical interlock, radiation monitors, source exposure indicator lights, viewing and intercom systems, and clock and decayed source activity in the unit's computer; and
(5) The name of the individual who performed the periodic spot-check and the signature of the authorized medical physicist who reviewed the record of the spot-check.
(c) A licensee shall retain a copy of the procedures required by § 35.643(b) until the licensee no longer possesses the remote afterloader unit.
Legislative history
This is a federal regulation, adopted through agency rulemaking under the Administrative Procedure Act — not enacted by a recorded vote of Congress.