10 CFR 35.2645
§ 35.2645 Records of periodic spot-checks for gamma stereotactic radiosurgery units.
United States · 10 CFR — Energy · Status: effective
Cite this
- Citation
- 10 CFR 35.2645, § 35.2645 Records of periodic spot-checks for gamma stereotactic radiosurgery units, United States, version 1 as recorded 2026-07-09, yourstate.us, https://yourstate.us/provision/37042
- Permanent ID
ys:prov:37042@1- SHA-256
49017395d8e9226367767d8cb4360c16b9a0ff1dfeb96be1a7bbe00c68ef1f17
The hash is SHA-256 of this version's text, with every run of whitespace collapsed to a single space and the ends trimmed. The ID always leads back here, and checking it says whether the text you cited is still the current version.
Full text
(a) A licensee shall retain a record of each spot-check for gamma stereotactic radiosurgery units required by § 35.645 for 3 years.
(b) The record must include—
(1) The date of the spot-check;
(2) The manufacturer's name, model number, and serial number for the gamma stereotactic radiosurgery unit and the instrument used to measure the output of the unit;
(3) An assessment of timer linearity and accuracy;
(4) The calculated on-off error;
(5) A determination of trunnion centricity;
(6) The difference between the anticipated output and the measured output;
(7) An assessment of source output against computer calculations;
(8) Notations indicating the operability of radiation monitors, helmet microswitches, emergency timing circuits, emergency off buttons, electrical interlocks, source exposure indicator lights, viewing and intercom systems, timer termination, treatment table retraction mechanism, and stereotactic frames and localizing devices (trunnions); and
(9) The name of the individual who performed the periodic spot-check and the signature of the authorized medical physicist who reviewed the record of the spot-check.
(c) A licensee shall retain a copy of the procedures required by § 35.645(b) until the licensee no longer possesses the gamma stereotactic radiosurgery unit.
Legislative history
This is a federal regulation, adopted through agency rulemaking under the Administrative Procedure Act — not enacted by a recorded vote of Congress.