10 CFR 35.2647
§ 35.2647 Records of additional technical requirements for mobile remote afterloader units.
United States · 10 CFR — Energy · Status: effective
Cite this
- Citation
- 10 CFR 35.2647, § 35.2647 Records of additional technical requirements for mobile remote afterloader units, United States, version 1 as recorded 2026-07-09, yourstate.us, https://yourstate.us/provision/37043
- Permanent ID
ys:prov:37043@1- SHA-256
4829858ec832c23953cf99c271c002344df8b5e98939fdd31ba54c0ab9f7283a
The hash is SHA-256 of this version's text, with every run of whitespace collapsed to a single space and the ends trimmed. The ID always leads back here, and checking it says whether the text you cited is still the current version.
Full text
(a) A licensee shall retain a record of each check for mobile remote afterloader units required by § 35.647 for 3 years.
(b) The record must include—
(1) The date of the check;
(2) The manufacturer's name, model number, and serial number of the remote afterloader unit;
(3) Notations accounting for all sources before the licensee departs from a facility;
(4) Notations indicating the operability of each entrance door electrical interlock, radiation monitors, source exposure indicator lights, viewing and intercom system, applicators, source transfer tubes, and transfer tube applicator interfaces, and source positioning accuracy; and
(5) The signature of the individual who performed the check.
Legislative history
This is a federal regulation, adopted through agency rulemaking under the Administrative Procedure Act — not enacted by a recorded vote of Congress.