§ 866.3361 Mass spectrometer system for clinical use for the identification of microorganisms.

United States · 21 CFR — Food and Drugs · Status: effective

Get this as JSONEmbed this
Cite this
Citation
21 CFR 866.3361, § 866.3361 Mass spectrometer system for clinical use for the identification of microorganisms, United States, version 1 as recorded 2026-07-09, yourstate.us, https://yourstate.us/provision/89532
Permanent ID
ys:prov:89532@1
SHA-256
6f1de9aeb44aac29ea483e4fbff066fbb55a63fb2df8975b6239f7b13c610239

The hash is SHA-256 of this version's text, with every run of whitespace collapsed to a single space and the ends trimmed. The ID always leads back here, and checking it says whether the text you cited is still the current version.

Full text

(a) Identification. A mass spectrometer system for clinical use for the identification of microorganisms is a qualitative in vitro diagnostic device intended for the identification of microorganisms cultured from human specimens. The device is comprised of an ionization source, a mass analyzer, and a spectral database. The device is indicated for use in conjunction with other clinical and laboratory findings to aid in the diagnosis of bacterial and fungal infections. (b) Classification. Class II (special controls). The special controls for this device are: (1) Premarket notification submissions must include detailed documentation for device software, including, but not limited to, standalone software applications and hardware-based devices that incorporate software. (2) Premarket notification submissions must include database implementation methodology, construction parameters, and quality assurance protocols. (3) A detailed explanation of the interpretation of results and acceptance criteria must be included in the device's 21 CFR 809.10(b)(9) compliant labeling. (4) As part of the risk management activities performed under 21 CFR 820.10(c) design and development, you must document an appropriate end user device training program that will be offered as part of your efforts to mitigate the risk of failure to correctly operate the instrument. (5) Premarket notification submissions must include details on the appropriate end user device training program that will be offered while marketing the device.

Legislative history

This is a federal regulation, adopted through agency rulemaking under the Administrative Procedure Act — not enacted by a recorded vote of Congress.