§ 870.1251 Temporary catheter for embolic protection during transcatheter intracardiac procedures.

United States · 21 CFR — Food and Drugs · Status: effective

Get this as JSONEmbed this
Cite this
Citation
21 CFR 870.1251, § 870.1251 Temporary catheter for embolic protection during transcatheter intracardiac procedures, United States, version 1 as recorded 2026-07-09, yourstate.us, https://yourstate.us/provision/89880
Permanent ID
ys:prov:89880@1
SHA-256
28d5a5187a8fa2d2c185d4393c7e3cc844ce37f6b852a3ecc66137f01fc20acd

The hash is SHA-256 of this version's text, with every run of whitespace collapsed to a single space and the ends trimmed. The ID always leads back here, and checking it says whether the text you cited is still the current version.

Full text

(a) Identification. This device is a single use percutaneous catheter system that has (a) blood filter(s) at the distal end. This device is indicated for use while performing transcatheter intracardiac procedures. The device is used to filter blood in a manner that may prevent embolic material (thrombus/debris) from the transcatheter intracardiac procedure from traveling towards the cerebral circulation. (b) Classification. Class II (special controls). The special controls for this device are: (1) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (i) Simulated-use testing in a clinically relevant bench anatomic model to assess the following: (A) Delivery, deployment, and retrieval, including quantifying deployment and retrieval forces, and procedural time; and (B) Device compatibility and lack of interference with the transcatheter intracardiac procedure and device. (ii) Tensile strengths of joints and components, tip flexibility, torque strength, torque response, and kink resistance. (iii) Flow characteristics. (A) The ability of the filter to not impede blood flow. (B) The amount of time the filter can be deployed in position and/or retrieved from its location without disrupting blood flow. (iv) Characterization and verification of all dimensions. (2) Animal testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be assessed: (i) Delivery, deployment, and retrieval, including quantifying procedural time. (ii) Device compatibility and lack of interference with the transcatheter intracardiac procedure and device. (iii) Flow characteristics. (A) The ability of the filter to not impede blood flow. (B) The amount of time the filter can be deployed in position and/or retrieved from its location without disrupting blood flow. (iv) Gross pathology and histopathology assessing vascular injury and downstream embolization. (3) All patient contacting components of the device must be demonstrated to be biocompatible. (4) Performance data must demonstrate the sterility of the device components intended to be provided sterile. (5) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life. (6) Labeling for the device must include: (i) Instructions for use; (ii) Compatible transcatheter intracardiac procedure devices; (iii) A detailed summary of the clinical testing conducted; and (iv) A shelf life and storage conditions. (7) Clinical performance testing must demonstrate: (i) The ability to safely deliver, deploy, and remove the device; (ii) The ability of the device to filter embolic material while not impeding blood flow; (iii) Secure positioning and stability of the position throughout the transcatheter intracardiac procedure; and (iv) Evaluation of all adverse events including death, stroke, and vascular injury.

Legislative history

This is a federal regulation, adopted through agency rulemaking under the Administrative Procedure Act — not enacted by a recorded vote of Congress.