21 CFR 874.3325
§ 874.3325 Self-fitting air-conduction hearing aid.
United States · 21 CFR — Food and Drugs · Status: effective
Cite this
- Citation
- 21 CFR 874.3325, § 874.3325 Self-fitting air-conduction hearing aid, United States, version 1 as recorded 2026-07-09, yourstate.us, https://yourstate.us/provision/90213
- Permanent ID
ys:prov:90213@1- SHA-256
07e4692a4fa04683b3b68c94efc280517e2c595a788d7be4284d22f9a0d04ae6
The hash is SHA-256 of this version's text, with every run of whitespace collapsed to a single space and the ends trimmed. The ID always leads back here, and checking it says whether the text you cited is still the current version.
Full text
(a) Identification. A self-fitting air-conduction hearing aid is a wearable sound amplifying device that is intended to compensate for impaired hearing and incorporates technology, including software, that allows users to program their hearing aids. This technology integrates user input with a self-fitting strategy and enables users to independently derive and customize their hearing aid fitting and settings. A self-fitting air-conduction hearing aid is subject to the requirements in § 800.30 or § 801.422 of this chapter, as applicable.
(b) Classification. Class II (special controls). The special controls for this device are:
(1) Clinical data must evaluate the effectiveness of the self-fitting strategy.
(2) Electroacoustic parameters, including maximum output limits, distortion levels, self-generated noise levels, latency, and frequency response, must be specified and tested.
(3) Performance data must demonstrate the electromagnetic compatibility (EMC), electrical safety, and thermal safety of the device.
(4) Software verification, validation, and hazard analysis must be performed.
(5) If the device incorporates wireless technology:
(i) Performance testing must validate safety of exposure to non-ionizing radiation; and
(ii) Performance data must validate wireless technology functions.
(6) Usability testing must demonstrate that users can correctly use the device as intended under anticipated conditions of use.
Legislative history
This is a federal regulation, adopted through agency rulemaking under the Administrative Procedure Act — not enacted by a recorded vote of Congress.