21 CFR 876.4350

§ 876.4350 Fluid jet system for prostate tissue removal.

United States · 21 CFR — Food and Drugs · Status: effective

Get this as JSONEmbed this
Cite this
Citation
21 CFR 876.4350, § 876.4350 Fluid jet system for prostate tissue removal, United States, version 1 as recorded 2026-07-09, yourstate.us, https://yourstate.us/provision/90302
Permanent ID
ys:prov:90302@1
SHA-256
e67758f53e10d94753bd971e7d00b87cca94d898d1fdb69faa0d42e7374015d0

The hash is SHA-256 of this version's text, with every run of whitespace collapsed to a single space and the ends trimmed. The ID always leads back here, and checking it says whether the text you cited is still the current version.

Full text

(a) Identification. A fluid jet system for prostate tissue removal is a prescription device intended for the resection and removal of prostatic tissue for the treatment of benign prostatic hyperplasia. The device cuts tissue by using a pressurized jet of fluid delivered to the prostatic urethra. The device is able to image the treatment area, or pairs with an imaging modality, to monitor treatment progress. (b) Classification. Class II (special controls). The special controls for this device are: (1) Clinical performance testing must evaluate the following: (i) All adverse events associated with the device, and (ii) Improvement in lower urinary tract symptoms (LUTS). (2) Physician training must be provided that includes: (i) Information on key aspects and use of the device, and (ii) Information on how to override or stop resection. (3) Animal testing must demonstrate that the device resects targeted tissue in a controlled manner without injury to adjacent non-target tissues. (4) Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested: (i) Measurement of targeting accuracy and reproducibility of high velocity fluid jet, and (ii) High pressure fluid jet verification testing at target and non-target tissues. (5) Software verification, validation, and hazard analysis must be performed. (6) The patient-contacting elements of the device must be demonstrated to be biocompatible. (7) Performance data must demonstrate the electrical safety and electromagnetic compatibility of the device. (8) Performance data must demonstrate the sterility of the patient-contacting components of the device. (9) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the identified shelf life. (10) Performance data must validate the instructions for reprocessing and reliability of reusable components. (11) Labeling must include the following: (i) A section that summarizes the clinical testing results, including the adverse event profile and improvement in LUTS; (ii) A shelf life for single use components; (iii) A use life for reusable components; and (iv) Reprocessing instructions for reusable components.

Legislative history

This is a federal regulation, adopted through agency rulemaking under the Administrative Procedure Act — not enacted by a recorded vote of Congress.