21 CFR 876.4410
§ 876.4410 Endoscopic traction device.
United States · 21 CFR — Food and Drugs · Status: effective
Cite this
- Citation
- 21 CFR 876.4410, § 876.4410 Endoscopic traction device, United States, version 1 as recorded 2026-07-09, yourstate.us, https://yourstate.us/provision/90305
- Permanent ID
ys:prov:90305@1- SHA-256
3d6ca80b32043db1142b7315c11566114a1b69017fc58f5a8148dd33a5367076
The hash is SHA-256 of this version's text, with every run of whitespace collapsed to a single space and the ends trimmed. The ID always leads back here, and checking it says whether the text you cited is still the current version.
Full text
(a) Identification. An endoscopic traction device is a prescription device that is endoscopically applied to retract tissue in the gastrointestinal tract during dissection procedures to increase visualization of the dissection plane and assist in tissue resection, exposure, and removal.
(b) Classification. Class II (special controls). The special controls for this device are:
(1) In vivo performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Testing must evaluate:
(i) Perforation, bleeding, and mucosal injury;
(ii) Ease of insertion and removal of the device;
(iii) Visualization during the procedure; and
(iv) Ease of procedure as reported by the intended user.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Testing must include:
(i) Device deployment and detachment;
(ii) Ability to retract tissue;
(iii) Tensile strength;
(iv) Potential for laceration caused by the device or procedure using the device;
(v) Dimensional verification; and
(vi) For devices that contain a magnet, magnet strength verification and safety assessment.
(3) Usability assessment must demonstrate that the intended user(s) can safely and correctly use the device.
(4) Performance data must demonstrate the sterility of the patient-contacting components of the device.
(5) The patient-contacting components of the device must be demonstrated to be biocompatible.
(6) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the intended shelf life.
(7) Labeling must include:
(i) The recommended training for safe use of the device;
(ii) Anatomical locations and lesion sizes that have been demonstrated to be safe to use with the device; and
(iii) A shelf life.
Legislative history
This is a federal regulation, adopted through agency rulemaking under the Administrative Procedure Act — not enacted by a recorded vote of Congress.