21 CFR 878.4685
§ 878.4685 Extracorporeal shock wave device for treatment of chronic wounds.
United States · 21 CFR — Food and Drugs · Status: effective
Cite this
- Citation
- 21 CFR 878.4685, § 878.4685 Extracorporeal shock wave device for treatment of chronic wounds, United States, version 1 as recorded 2026-07-09, yourstate.us, https://yourstate.us/provision/90451
- Permanent ID
ys:prov:90451@1- SHA-256
3e65cdee5c3c6824d0ec776a1629a6edcba5aa799a0230f24e805c8cc1015bd6
The hash is SHA-256 of this version's text, with every run of whitespace collapsed to a single space and the ends trimmed. The ID always leads back here, and checking it says whether the text you cited is still the current version.
Full text
(a) Identification. An extracorporeal shock wave device for treatment of chronic wounds is a prescription device that focuses acoustic shock waves onto the dermal tissue. The shock waves are generated inside the device and transferred to the body using an acoustic interface.
(b) Classification. Class II (special controls). The special controls for this device are:
(1) Non-clinical performance testing must be conducted to demonstrate that the system produces anticipated and reproducible acoustic pressure shock waves.
(2) The patient-contacting components of the device must be demonstrated to be biocompatible.
(3) Performance data must demonstrate that the reusable components of the device can be reprocessed for subsequent use.
(4) Performance data must be provided to demonstrate the electromagnetic compatibility and electrical safety of the device.
(5) Software verification, validation, and hazard analysis must be performed.
(6) Performance data must support the use life of the system by demonstrating continued system functionality over the labeled use life.
(7) Physician labeling must include:
(i) Information on how the device operates and the typical course of treatment;
(ii) A detailed summary of the device's technical parameters;
(iii) Validated methods and instructions for reprocessing of any reusable components; and
(iv) Instructions for preventing hearing loss by use of hearing protection.
(8) Patient labeling must include:
(i) Relevant contraindications, warnings, precautions, adverse effects, and complications;
(ii) Information on how the device operates and the typical course of treatment;
(iii) The probable risks and benefits associated with the use of the device;
(iv) Post-procedure care instructions; and
(v) Alternative treatments.
Legislative history
This is a federal regulation, adopted through agency rulemaking under the Administrative Procedure Act — not enacted by a recorded vote of Congress.