21 CFR 880.6305
§ 880.6305 Ingestible event marker.
United States · 21 CFR — Food and Drugs · Status: effective
Cite this
- Citation
- 21 CFR 880.6305, § 880.6305 Ingestible event marker, United States, version 1 as recorded 2026-07-09, yourstate.us, https://yourstate.us/provision/90585
- Permanent ID
ys:prov:90585@1- SHA-256
d92a9efe43d2a4de1edcafc3990421a21a6779380dd9ad4d05714f60f3e46d5b
The hash is SHA-256 of this version's text, with every run of whitespace collapsed to a single space and the ends trimmed. The ID always leads back here, and checking it says whether the text you cited is still the current version.
Full text
(a) Identification. An ingestible event marker is a prescription device used to record time-stamped, patient-logged events. The ingestible component links wirelessly through intrabody communication to an external recorder which records the date and time of ingestion as well as the unique serial number of the ingestible device.
(b) Classification. Class II (special controls). The special controls for this device are:
(1) The device must be demonstrated to be biocompatible and non-toxic;
(2) Nonclinical, animal, and clinical testing must provide a reasonable assurance of safety and effectiveness, including device performance, durability, compatibility, usability (human factors testing), event recording, and proper excretion of the device;
(3) Appropriate analysis and nonclinical testing must validate electromagnetic compatibility performance, wireless performance, and electrical safety; and
(4) Labeling must include a detailed summary of the nonclinical and clinical testing pertinent to use of the device and the maximum number of daily device ingestions.
Legislative history
This is a federal regulation, adopted through agency rulemaking under the Administrative Procedure Act — not enacted by a recorded vote of Congress.