§ 880.6600 Ultraviolet (UV) radiation chamber disinfection device.

United States · 21 CFR — Food and Drugs · Status: effective

Get this as JSONEmbed this
Cite this
Citation
21 CFR 880.6600, § 880.6600 Ultraviolet (UV) radiation chamber disinfection device, United States, version 1 as recorded 2026-07-09, yourstate.us, https://yourstate.us/provision/90594
Permanent ID
ys:prov:90594@1
SHA-256
5db67b01f377a56b987c15af37f120ac351af3f9535df2be515c2f9812cc3ccb

The hash is SHA-256 of this version's text, with every run of whitespace collapsed to a single space and the ends trimmed. The ID always leads back here, and checking it says whether the text you cited is still the current version.

Full text

(a) Identification. An ultraviolet (UV) radiation chamber disinfection device is intended for the low-level surface disinfection of non-porous equipment surfaces by dose-controlled UV irradiation. This classification does not include self-contained open chamber UV radiation disinfection devices intended for whole room disinfection in a health care environment. (b) Classification—Class II (special controls). The special controls for this device are: (1) Performance testing must demonstrate the following: (i) The chamber's ability to control the UV radiation dose during operation. (ii) The chamber's disinfection performance through microbial challenge testing. (iii) Evidence that the equipment intended to be processed is UV compatible. (iv) Validation of the cleaning and disinfection procedures. (v) The ability of the device to continue to perform to all specification after cleaning and disinfection. (vi) Whether the device generates ozone (if so, 21 CFR 801.415, Maximum acceptable level of ozone, applies). (2) Appropriate software verification, validation, and hazard analysis must be performed. (3) Appropriate analysis and/or testing must validate electrical safety, mechanical safety, and electromagnetic compatibility of the device in its intended use environment. (4) The labeling must include: (i) UV hazard warning labels. (ii) Explanation of all displays and/or labeling on user interface. (iii) Explanation of device safety interlocks. (iv) Explanation of all disinfection cycle signals, cautions and warnings. (v) Device operating procedures. (vi) Identification of the expected UV lamp operational life and instructions for procedures on replacement of the UV lamp when needed. (vii) Procedures to follow in case of UV lamp malfunction or failure. (viii) Procedures for disposing of mercury-containing UV lamps, if applicable. (ix) Identification of specific equipment that is compatible with the UV radiation dose generated by the device and that can safely undergo UV radiation low-level disinfection in the chamber device. (x) Description of the required preparation of equipment for disinfection in the UV radiation chamber device. (xi) Identification of the specific microbes used in successful performance testing of the device. (xii) Validated instructions for cleaning and disinfection of the device.

Legislative history

This is a federal regulation, adopted through agency rulemaking under the Administrative Procedure Act — not enacted by a recorded vote of Congress.