§ 884.4910 Specialized surgical instrumentation for use with urogynecologic surgical mesh.

United States · 21 CFR — Food and Drugs · Status: effective

Get this as JSONEmbed this
Cite this
Citation
21 CFR 884.4910, § 884.4910 Specialized surgical instrumentation for use with urogynecologic surgical mesh, United States, version 1 as recorded 2026-07-09, yourstate.us, https://yourstate.us/provision/90837
Permanent ID
ys:prov:90837@1
SHA-256
c1218f4f9afa1de8912bd565f699f335bd133b69a5a0b2e5d9d33974f086bf53

The hash is SHA-256 of this version's text, with every run of whitespace collapsed to a single space and the ends trimmed. The ID always leads back here, and checking it says whether the text you cited is still the current version.

Full text

(a) Identification. Specialized surgical instrumentation for use with urogynecologic surgical mesh is a prescription device specifically intended for use as an aid in the insertion, placement, fixation, or anchoring of surgical mesh during urogynecologic procedures. These procedures include transvaginal pelvic organ prolapse repair, sacrocolpopexy (transabdominal pelvic organ prolapse repair), and treatment of female stress urinary incontinence. Examples of specialized surgical instrumentation include needle passers and trocars, needle guides, fixation tools, and tissue anchors. This device is not a manual gastroenterology-urology surgical instrument and accessories (§ 876.4730) or a manual surgical instrument for general use (§ 878.4800). (b) Classification. Class II (special controls). The special controls for specialized surgical instrumentation for use with urogynecologic surgical mesh are: (1) The device must be demonstrated to be biocompatible; (2) The device must be demonstrated to be sterile and, if reusable, it must be demonstrated that the device can be adequately reprocessed; (3) Performance data must support the shelf life of the device by demonstrating package integrity and device functionality over the requested shelf life; (4) Non-clinical performance testing must demonstrate that the device meets all design specifications and performance requirements, and that the device performs as intended under anticipated conditions of use; and (5) Labeling must include: (i) Information regarding the mesh design that may be used with the device; (ii) Detailed summary of the clinical evaluations pertinent to use of the device; (iii) Expiration date; and (iv) Where components are intended to be sterilized by the user prior to initial use and/or are reusable, validated methods and instructions for sterilization and/or reprocessing of any reusable components.

Legislative history

This is a federal regulation, adopted through agency rulemaking under the Administrative Procedure Act — not enacted by a recorded vote of Congress.