§ 886.5201 Intense pulsed light device for managing dry eye.
United States · 21 CFR — Food and Drugs · Status: effective
Cite this
- Citation
- 21 CFR 886.5201, § 886.5201 Intense pulsed light device for managing dry eye, United States, version 1 as recorded 2026-07-09, yourstate.us, https://yourstate.us/provision/91008
- Permanent ID
ys:prov:91008@1- SHA-256
a42cb748f4952e8c323bd9d9a2b8c0911ee6bd35ec91e5f516df3c8e3583fd8d
The hash is SHA-256 of this version's text, with every run of whitespace collapsed to a single space and the ends trimmed. The ID always leads back here, and checking it says whether the text you cited is still the current version.
Full text
(a) Identification. An intense pulsed light device for managing dry eye is a prescription device intended for use in the application of intense pulsed light therapy to the skin. The device is used in patients with dry eye disease due to meibomian gland dysfunction, also known as evaporative dry eye or lipid deficiency dry eye.
(b) Classification. Class II (special controls). The special controls for this device are:
(1) Clinical performance testing must evaluate adverse events and improvement of dry eye signs and symptoms under anticipated conditions of use.
(2) Thermal safety assessment in a worst-case scenario must be performed to validate temperature safeguards.
(3) Performance testing must demonstrate electrical safety and electromagnetic compatibility (EMC) of the device in the intended use environment.
(4) Software verification, validation, and hazard analysis must be performed.
(5) The patient-contacting components of the device must be demonstrated to be biocompatible.
(6) Physician and patient labeling must include:
(i) Device technical parameters;
(ii) A summary of the clinical performance testing conducted with the device;
(iii) A description of the intended treatment area location;
(iv) Warnings and instructions regarding the use of safety-protective eyewear for patient and device operator;
(v) A description of intense pulse light (IPL) radiation hazards and protection for patient and operator;
(vi) Instructions for use, including an explanation of all user interface components; and
(vii) Instructions on how to clean and maintain the device and its components.