21 CFR 886.5310
§ 886.5310 Intranasal electrostimulation device for dry eye symptoms.
United States · 21 CFR — Food and Drugs · Status: effective
Cite this
- Citation
- 21 CFR 886.5310, § 886.5310 Intranasal electrostimulation device for dry eye symptoms, United States, version 1 as recorded 2026-07-09, yourstate.us, https://yourstate.us/provision/91011
- Permanent ID
ys:prov:91011@1- SHA-256
a7913da594a9f4140e5c05a29ae0f02682dfcc11d07e3bac223333e80eb7bf13
The hash is SHA-256 of this version's text, with every run of whitespace collapsed to a single space and the ends trimmed. The ID always leads back here, and checking it says whether the text you cited is still the current version.
Full text
(a) Identification. An intranasal electrostimulation device for dry eye symptoms is a prescription non-implantable, electrostimulation device intended to increase tear production for improvement in dry eye symptoms.
(b) Classification. Class II (special controls). The special controls for this device are:
(1) Clinical performance testing must evaluate improvement of dry eye symptoms under anticipated conditions of use.
(2) Non-clinical performance testing must assess the following electrical output specifications: waveforms, output modes, maximum output voltage, maximum output current, pulse duration, frequency, net charge per pulse, maximum phase charge at 500 ohms, maximum current density, maximum average current, and maximum average power density.
(3) Patient-contacting components of the device must be demonstrated to be biocompatible.
(4) Performance testing must demonstrate the electrical, thermal, and mechanical safety along with electromagnetic compatibility (EMC) of the device in the intended use environment.
(5) Software verification, validation, and hazard analysis must be performed.
(6) Training for the proper use of the device must be provided.
(7) Physician and patient labeling must include:
(i) Summaries of electrical stimulation parameters;
(ii) Instructions on how to correctly use and maintain the device;
(iii) Instructions and explanations of all user-interface components;
(iv) Information related to electromagnetic compatibility classification;
(v) Instructions on how to clean the device; and
(vi) Summaries of clinical performance testing demonstrating safety and effectiveness.